MINIMED™ 770G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
START HERE
1 2 3 4
Wash your hands. Select Reservoir & Tubing. Select New Reservoir. Remove the infusion set you have been using by
Press . Select Options. loosening the adhesive and pulling away from body

5 6
Remove the used reservoir from the pump. Select Rewind

FILL RESERVOIR & CONNECT TO THE INFUSION SET TUBING
Follow the next steps to fill reservoir with insulin and connect to the infusion set tubing

1 Plunger 2 3 Transfer
4
rod
guard
Reservoir
Transfer Plunger Insulin
guard vial
Remove from package. Make sure Pull plunger down to Wipe vial with alcohol. Place Push and hold
insulin vial is at room temperature the amount that you vial on table. Firmly press the plunger down

to reduce the risk of air bubbles. plan to fill with insulin. blue transfer guard onto vial

1
Continued on next page
MINIMED™ 770G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
WARNING If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly fill
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycaemia or hypoglycaemia

5 6 7 8
With your thumb still on the plunger, Tap the reservoir to move If needed, pull plunger To avoid getting insulin on the top of the
flip over so vial is on top. Release air bubbles to top of back down to amount reservoir, turn vial over so it is upright. Hold
thumb and pull plunger down to fill reservoir. Push plunger of insulin needed for transfer guard and turn reservoir counter-
with insulin. up to move air into vial. 2-3 days. clockwise and remove from transfer guard

CONNECT RESERVOIR TO INFUSION SET
You will place the tubing connector onto the end of the infusion set to the filled reservoir

1 Quick-serter® Tubing connector
2 3 4
Release Quick-set ®
button
Needle guard
Green side Green side
button button
Blue hub Hub handle
Tubing
slot Gently push the tubing If you see air bubbles, tap Twist plunger counter-
Remove infusion set from package. Remove connector onto reservoir. reservoir to move them to clockwise to loosen
the paper that holds the tubing together. Turn clockwise until locked. top. Push plunger just a bit to and remove

You will hear a click. move them into tubing

THE BACKLIGHT MAY
HAVE TURNED OFF
Press any button to turn
the screen back on

Press to open the menu. If the pump Select Load Reservoir Select Next

is locked, you will need to unlock the from the menu. 2
pump after pressing . Continued on next page
MINIMED™ 770G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
PLACE RESERVOIR INTO PUMP
Now place the filled reservoir into the reservoir compartment of the pump

1 2 3
Place reservoir into pump. Turn clockwise until you feel Select Next

reservoir lock into place

LOAD RESERVOIR AND FILL TUBING
Follow these steps to load the reservoir and fill the tubing

1 2 3 4
Drops at end
of tubing
Select Load and keep When you see this screen, Select Fill and keep holding until you see After you see drops,
holding . select Next. drops at the end of tubing, then let go. press and select Next

INSERT INFUSION SET
Next, follow the steps to insert the infusion set into your body

1 2 3
Place the MiniMed Quick-serter Use two fingers to seat the Peel the paper from the
insertion device onto a sturdy flat infusion set inside the serter adhesive on both sides
surface with handle facing down. securely and gently push down. of the needle guard

Place blue hub into serter, placing 3
the handle in the tubing slot. Continued on next page
MINIMED™ 770G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
4 5 6
Pull green handle down The best areas of the body to insert the infusion set are Turn to loosen needle
until it clicks. The two shown in the shaded areas of this drawing. Choose areas guard and pull

green side buttons will that are away from the natural bend of your waist or belt
also click into place. Do line of your pants. Avoid areas of hard or scar tissue. Consult
not press the green side with your healthcare provider on appropriate areas to insert
buttons at this time. your infusion set. Wipe with alcohol or other antiseptic

7 8 9 10
Hold serter against Press the two green side Press down the release Pull serter away from body. Press
cleaned site. buttons at the same time. button of the serter to adhesive against skin

If they are not pressed at the release the infusion set

same time the infusion set
will not insert properly

11 12 1
Dispose blue
handle into
a sharps
2 container

Hold infusion set. Pull blue handle straight out Fold blue handle
to remove needle. until locked

4
Continued on next page
MINIMED™ 770G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
FILL CANNULA
You will now fill the cannula, the little tube under your skin, with insulin

1 2
NOTE: Your pump will remember the Fill amount that you
used last. Always verify that the Fill amount is correct

ƒ If it is correct, press to Fill Now and press

ƒ If it is incorrect, press . Change to correct amount and
press . Select Fill Now

Select Fill. Select Fill amount and enter :
ƒ 0.300 if using 6mm cannula
ƒ 0.500 if using 9mm cannula
Then press

3 4
NOTE: Select Stop Filling if you need
to stop, for example, if you notice the
Total amount is incorrect. This should
rarely happen if you have verified the
Fill amount on the previous screen

Select Fill Now. The Home screen displays the
insulin as it fills the cannula

Your infusion set change is now complete!
Always read instructions for use
©2020 Medtronic MiniMed, Inc. All rights reserved. Medtronic Australasia Pty Ltd. 2 Alma Road, Macquarie Park. NSW 2113. Australia. medtronic-diabetes.com.au 8339-082020 5

0.300 if using 6mm cannula 0.500 if using 9mm cannula Then press . The Home screen displays the insulin as it fills the cannula. Your infusion set change is now complete! Medtronic …

Download Now

Documemt Updated

Popular Download

Frequently Asked Questions

What is the minimed quick set infusion set?

The MiniMed™ Quick-set™ infusion set provides the convenience of a disconnect at the infusion site and a pre-attached adhesive for a discreet, low profile design. Warning! Please Try Again Later.

What is the difference between the minimed quick serter and minimed silhouette device?

The MiniMed ™ Quick-serter ™ device is the one used with the MiniMed ™ Quick-set ™ infusion set and helps to make consistent insertions with just the press of a button. The MiniMed ™ Silhouette ™ infusion set is a soft cannula infusion set that can be inserted at a variety of angles.

What is an insulin infusion set?

Designed for Every Body and Every Lifestyle. An infusion set includes the thin plastic tubing that delivers insulin from the pump to your body. The MiniMed Quick-set infusion set is our most popular infusion set. The MiniMed Quick-set is available in four tubing lengths and two cannula lengths.

How many steps does it take to insert a minimed infusion?

This infusion set has a 90°, soft cannula and can be inserted with only 4 steps. The all-in-one insertion design means that no separate serter is needed. The MiniMed ™ Mio ™ Advance infusion set can be inserted with one hand (left or right) and has a hidden needle throughout insertion.